Bacteriostatic Water

Lot BACW-20260611-AB26 Tested October 7, 2026

Overall Result

Outside spec

Accumark Labs reported 3 results outside specification:

  • Benzyl alcohol measured at 0%, which is below the acceptable range of 0.72% – 1.08%.
  • Endotoxin measured at 19.7 EU/mL, which exceeds the limit of 0.25 EU/mL.
  • pH measured at 8.04, which exceeds the acceptable range of 4.5 – 7.

Test information

Lot number
BACW-20260611-AB26
Test date
October 7, 2026
Laboratory
Accumark Labs
Fill volume
Measured: 0.00 mL
Manufacturer
Pending

Analytical Results

Individual test results from the independent laboratory report.

Benzyl Alcohol

Measured 0%

Specification 0.9% (v/v) ±20%

Outside spec Below the lower limit of 0.72%

Endotoxin

Measured 19.7 EU/mL

Specification ≤0.25 EU/mL

Outside spec Above the upper limit of 0.25 EU/mL

Sterility

Result Not Detected

Method Rapid Sterility Screening (PCR)

Met spec

pH

Measured 8.04

Specification 4.5 – 7.0

Outside spec Above the upper limit of 7

Testing Methods View methods used for this report
Endotoxin
Endotoxin testing performed using Limulus Amebocyte Lysate assay in accordance with USP <85> under validated laboratory conditions.
Sterility
Rapid sterility screening performed by real-time polymerase chain reaction (qPCR) targeting conserved microbial DNA sequences under validated laboratory conditions.

HPLC Chromatogram

Chromatographic analysis of benzyl alcohol content (Accumark Labs).

Plotted from Accumark Labs' instrument data. Drag across the chart to zoom in; double-click to reset. Tap the chart to read a point.

Testing Summary

Summary of results, specifications, and laboratory verdicts.

Analytical results against the specifications entered for this report
Test Result Specification Status
Benzyl Alcohol 0% 0.9% (v/v) ±20% Outside spec
Endotoxin 19.7 EU/mL ≤0.25 EU/mL Outside spec
Sterility Not Detected Rapid Sterility Screening (PCR) Met spec
pH 8.04 4.5 – 7.0 Outside spec

Laboratory Documentation

Source information and report details from the testing laboratory.

Laboratory
Accumark Labs
Report number
AJK2-USQV
Received
October 5, 2026
Test date
October 7, 2026
Published
October 16, 2026
Last updated
October 16, 2026

View Laboratory COA (opens in a new tab)